How to use NanoCuvette™ One with SpectroWorks™?

Here you can read more about the steps needed to use SpectroWorks™ with NanoCuvette™ One.

Download PDF here.

 

You can read about getting started with NanoCuvette™ One here

 Prepare for measurement 

 

EXTF_20210527_Spectroworks_prepare_for_measurement

Sign in to app.spectroworks.com

 

Read how to create an account for SpectroWorks™ here.

 

PHOT_20200512_SpectroWorks_login (1)

 

 

 


Step 1 - Start

Step 1.1: Create project and measurement

 

Create a new project and choose your desired results

 

PHOT_20200512_SpectroWorks_create_new_project

 

Go to the Create tab to start a measurement 

 

    PHOT_20200512_SpectroWorks_tabs

 

 

 


Step 1 - Start

Step 1.2: Enter box code

 

Find the box code at the bottom of your NanoCuvette™ box

 

  EXTF_20210527_Spectroworks_enter_box_code-1

 

 

 


Step 1 - Start

Step 1.3: Select NanoCuvette™ in box

 

Pick a NanoCuvette™ in the box and select it by clicking the corresponding number on the screen. 

 

  EXTF_20210527_Spectroworks_pick_a_cuvette_number

 

 

 


Step 2 - Reference

Step 2.1: Acquire reference B-side spectrum

 


  • Carefully dispense DI- water into the NanoCuvette™.
  • Place the NanoCuvette™ in the cuvette holder of your spectrophotometer with the B- side (side with the optical filter) facing the light source. 
  • Acquire the reference B- side spectrum with your spectrophotometer. The scanning interval should include 550 - 800 nm. For best results use the highest possible scanning resolution. The software supports up to 4500 data points per spectrum. 
  • Save the measurements as .txt or .csv file. 

 

 


Step 2 - Reference

2.2: Upload reference spectrum

 

  • Upload your reference spectrum by dragging the file onto the blue box or clicking on it and choosing your file. 
  • Once uploaded, the reference spectrum for DI- water will be shown. Click "Next" to fit the software model to your reference spectrum.  
  EXTF_20210527_Spectroworks_01_upload_reference_spectrum

 

 

 


Step 2 - Reference

Step 2.3: Check reference quality

 

  • The reference spectrum is acceptable if the fit quality is above 75%. If the quality is lower, try the following and then acquire a new reference spectrum:
  • Ensure that the NanoCuvette™ is placed in the instrument with the optical filter facing the light source
  • Ensure that the height of the light path of your spectrophotometer is between 8.5 mm - 15 mm.
  • Try another unused 


  EXTF_20190115_Fitting_complete_Air_sample

 

 

 


Step 3 - Sample

3.1: Acquire sample B- side spectrum

 

undefined-3

Remember to acquire the spectrum using the same settings as in step 3

 

Dispose the DI- water from the NanoCuvette™ and carefully dispense your sample into the NanoCuvette™. Acquire the spectrum through your NanoCuvette™ with the B- side facing the light source. Save and upload the spectrum. Click "Next" to fit to the software model.

 

 

 

 

 


Step 3 - Sample

3.2 Acquire sample A- side spectrum

 

EXTF_20210527_Spectroworks_01_NC_A_side_spectrum

Keep your sample in your NanoCuvette™ 90 degrees such that the optical filter is not in the light path. Acquire the spectrum through your NanoCuvette™ with the A- side facing  the light source. Save and upload the spectrum. 

 

 

 

 

 

 

 


Step 4 - Finish

Step 4.1: View and export results 

 

View your experiment setup, spectra and results. It is also possible for you to type in your sample attributes and  notes, for you to easily compare experiments and print a report. The spectra can be exported by clicking the "Export" button. 

 

EXTF_20200304_Spectroworks_01_cut_to_frame

 

 

 

 

 